Prior to disassembling the system, we recommend that you read the processor's manual.
Disassembly instructions:
To begin, please use the picture below and get familiar with the system's main components.
1. Make sure the ultrasonic processor is turned off;
2. Detach the ultrasound cable from the transducer;
3. Make sure there is no liquid in the main liquid hoses;
4. Disassemble the inlet and outlet connections by removing the 1/2" clamps, gaskets and hoses;
5. Remove the 1.5” clamp holding the FBH-type Barbell horn (also called probe or sonotrode) in the reactor chamber;
6. Raise the ultrasonic stack higher on the support stand;
7. Remove the reactor chamber (also called flow cell) from the support stand;
8. Disassemble the support arm from the reactor chamber by using an adjustable wrench;
9. Remove the ultrasonic stack from the support stand;
10. Remove the transducer’s support arm by using an adjustable wrench;
12. Insert the pin of the second spanner wrench in a hole of the horn’s input end and set it up to provide counterclockwise torque;
14. Next, remove the Mylar acoustic washer to make sure that it is clean of any particles or scratches.
The LSP-500 ultrasonic processor is now disassembled!
If you need additional instructions, please download the processors's manual. For assembly instructions, please read our recent blog: Step-by-step Guide to Assembling Your LSP-500 Ultrasonic Processor.
To accompany these instructions, please see our video on LSP-500 Ultrasonic Processor Assembly and Disassembly procedures: